Mixer scale calibration

Your mixer scale drifts and nobody notices. The drift does not lock up, beep or light a warning — the scale simply starts weighing wrong, and the error contaminates the ration, the reports and your cost of gain.

A TMR mixer works under conditions no commercial scale would survive: constant vibration, loading impact, dust, water and temperature swings through the day. Load cells take all of it — and the drift they accumulate gives no warning.

Because the ration is formulated on that weight, the error spreads across the whole operation: at the feed mill, in the mix, in delivery and at the bunk.

Why this actually costs money

Feed is the largest cost in a feedlot. Every control in the operation — feed conversion, cost of gain, projected performance — assumes the recorded weight is the real weight.

The worst case is not the scale breaking. It is the scale appearing to work fine, drifting silently, for months.

How often to calibrate

Every 3 months is the right interval.

That cadence keeps drift under control in normal feedlot routine. But the calendar is not the only rule — bring the check forward whenever there is:

  • A load cell replacement
  • An impact or collision involving the mixer
  • Structural maintenance
  • Any sign of inconsistent readings

Signs the scale needs calibration

A single one of these justifies a check, even if the 3 months are not up yet.

What BULLTECN does during calibration

The service has three stages. None of them works alone — it is the combination that restores confidence in the weight.

1. Setup matched to the load cell model

Every load cell model has its own characteristics. The indicator must be configured for the cell actually installed on the mixer — beyond the weight unit, display increment and maximum capacity. Calibrating on a wrong setup means calibrating on a wrong basis: the work will have to be redone.

2. Data transmission speed from cell to indicator

The load cell sends readings to the indicator, and the speed of that transmission has to be matched to the mixer type and the way the operation uses it. The wrong setting produces unstable readings, drift and slow weight settling during loading — problems easily mistaken for a faulty scale, which they are not.

3. Calibration verified against a known weight

Verification is done against a known reference weight. And here is a clarification that saves the feedlot money:

You do not need to buy a certified test weight.

Verification can be done with what the feedlot already has in stock — bagged supplement of known weight, for instance. What determines calibration quality is not the cost of the reference, it is the correct procedure.

How the CAL factor works

CAL is the calibration factor stored in the scale indicator. Once setup is correct, the adjustment follows this logic:

  1. Place the known reference weight in the mixer
  2. Note the weight shown on the display
  3. Note the indicator's current CAL
  4. Calculate the new factor
  5. Write the new CAL to the indicator
new CAL = (known weight × current CAL) ÷ weight shown on the display

Always confirm with a second weighing. After writing the new CAL, weigh the reference again. If the reading still does not match, the problem is not the calibration factor — it is the load cell, cabling, connector or mounting. At that point it is repair, not calibration.

We service any feedlot

You do not need to have bought equipment from us. We provide calibration and field service in feedlots across Brazil, regardless of where the equipment was purchased.

Request an assessment

If your scale shows any of the signs above, or if it has been more than 3 months since the last check, get in touch.

FAQ

Questions about calibration

How often should the mixer scale be calibrated?

Every 3 months. Bring it forward whenever a load cell is replaced, the mixer takes an impact, structural maintenance is done, or readings look inconsistent.

Do I need a certified test weight?

Not necessarily. Verification can be done with a known weight the feedlot already has in stock, such as bagged supplement. What determines quality is the correct procedure: setup matched to the load cell model, transmission speed configured for the mixer type, and verification after writing the new CAL.

What is the CAL factor?

It is the calibration factor stored in the indicator. To adjust it:

new CAL = (known weight × current CAL) ÷ weight shown on the display